Description :
Tenders for Supply Of Ftir Spectrometer , Should Collect Ir Spectrum In Different Modes Such As Transmittance, Absorbance Pellets, Stand- Alone Films And Atr Mode , Spectral Range- Mid Ir - 7800 To 350 Cm-1 , Spectral Resolution- At Least 0.6 Cm-1 Or Lower In Mid Ir Range Without Using Any External J-stop Card , Wavenumber Precision- Lower Than 0.002 Cm-1 At 2000 Cm- 1 Or Better , Wavenumber Accuracy- Lower Than 0.06 Cm-1 At 2000 Cm-1 Or Better , Signal To Noise Ratio , Beam Splitter , Detectors , Atr Assembly- Single-reflection Atr With Diamond , Calibration Standards And Performance Verification , Sample Holder , Spectral Library , Software For Data Acquisition, Processing, And Evaluation. , Pc Requirement , Work Bench For The System , Warranty