Description :
Tenders for Supply Of Srphc153055-coronary Drug Eluting Stents With (1) Stent On Stainless Steel Or Cobalt Chromium Alloy Or Platinum Chromium Alloy. (2) Stent With Coating Of Polymer Carrier For Active Drug ( Sirolimus Everolimus Paclitaxel Or Zotarolimus Shown To Be Effective In Preventing Neo- Intimal Hyperplasia) (3) Stent Diameter Should Be From 2.5mm To 4mm And Length Of The Stent From 12mm To 38 Mm. (4) The Strut Thickness Should Be 91 Micron Or Less. (5) Stent Delivery System Shou Ld Be Monorail (single Operator Exchange Or Rapid Exchange). (6) Stent Should Be Compatible With 5fr Guiding Catheter (7) Stent Design Should Be Open Cell Or Hybrid Cell Type (8) Stent Should Be A Pproved By Dcgi. The Copy Of Dcgi Approval Should Be Enclosed In The Tender Bid. (9) The Long Term Efficacy For Ten Years And Safety Of The Stent In Terms Of Target Vessel Revascularisation (tvr) A Nd In-situ Thrombosis (ist) Should Be Documented / Published In Peer Reviewed Scientific Journals.( 10) The Documents Pertaining To The Efficacy And Safety Of The Stent Should Be Enclosed For The Brand That Is Offered In The Tender Bid (11) The Manufacturer/supplier Should Submit Proof Of Sup Ply Of The Stent To A Major Government Institute. Unit:no [quantity Tolerance (+/-): 5 %age , Item Category : Normal , Total Po Value Variation Permitted: Max 8 La Cs ]