Description :
Tenders for Srphc153053-coronary Drug Eluting Stents (1) With Single Wire Design With Cobalt All Oy Shell And Platinum Iridium Core (2) Stent With Coating Of Biolinx Polymer Carrier For Active Dr Ug Zotarolimus Shown To Be Effective In Preventing Neo-intimal Hyperplasia (3) Stent Diameter Sho Uld Be From 2 Mm To 5mm And Length Of The Stent From 8 Mm To 38 Mm. (4) The Strut Should Be Roun D With Thickness Ranging From 81 To 91 Microns (5) Stent Delivery System Should Be Monorail (sing Le Operator Exchange Or Rapid Exchange). (6) Stent Should Be Compatible With 5fr Guiding Cathete R (7) Stent Design Should Be Open Cell Type. (8) Stent Should Be Approved By Dcgi. The Copy Of Dcgi Approval Should Be Enclosed In The Tender Bid. (9) Five Years Efficacy And Safety Of The Stent In Te Rms Of Target Vessel Revascularisation (tvr) And In-situ Thrombosis (ist) Should Be Documented / Published In Peer Reviewed Scientific Journals.(10) The Documents Pertaining To The Efficacy And S Afety Of The Stent Should Be Enclosed For The Brand That Is Offered In The Tender Bid (11) The Ma Nufacturer/supplier Should Submit Proof Of Supply Of The Stent To A Major Government Institute. Unit:no [quantity Tolerance (+/-): 5 %age , Item Category : Normal , Total Po Value Variation Permitted: Max 8 La Cs ]