Description :
Tenders for Supply Of Srphc82278027-crt- P Pulse Generator For Delivery Of Cardiac Resynchronization Therapy. System Should Include A Current Generation Crt Pulse Generator Compatible With Bipolar Steroid-eluting Atrial And Ventricular Leads And A Biploar Left Ventricular Lead (coronary Sinus ). Should Provide Independently Programmable Right And Left Ventricular Pulse Outputs And Interventricular Delay ( With The Choices Of Right Ventricular Alone, Leftventricular Alone Or Biventricular Pacing As Well As The Choice Of Which Ventricle Is To Bepaced First With Optimization Of Delay For Pacing Of Other Ventricle ). Should Have Features Like Negative Av Hysteresis And Triggered Modes For Continuous Delivery Of Crt. Should Have Advanced Features For Detection And Management And Prevention Of Atrialtachycardias And Atrial Fibrillation As Well As Detailed Data Storage Of All Arrhythmias And Therapy As Well As Trends Over Time. Should Have Features For Management Of Atrial Fibrillation Including Mode Switch, Ventricular Rate Management Etc., Pulse Generator Should Be Suitable For Use With Either An Endocardial Lv Lead Or An Epimyocardial Lead For Lv Stimulation. Lv Leads Should Have Optimal Features For Introduction, Positioning And Fixation And Documented Low Displacement Rates. Safety And Efficacy Of The System Should Be Well Documented. Unit: Number. ]