Tender For corrigendum : supply of resuscitation equipment- made, 3 provided with closed ended reservoir with two valves, 4, patient valves pliable, well sealed, have minimum dead space and no forward or, backward leaks, 5, the bag should have an oxygen inlet which fits into the standard oxygen tubing both from, a cylinder and central supply, 6, face masks should be transparent, fit the patient outlet easily and have minimum dead, space., 7 the system should withstand washing scrubbing and autoclaving procedures, 8 face masks : 3 sizes i.e. 00,01,02,: 3 set with each bag., 9 should have option to control peep & pip., b self inflating bags 500 ml, 1 self inflating bag, 2 single/double walled silicone made, 3 provided with closed ended reservoir with two valves, 4, patient valves pliable, well sealed, have minimum dead space and no forward or, backward leaks, 5, the bag should have an oxygen inlet which fits into the standard oxygen tubing both from, a cylinder and central supply, 6, face masks should be transparent, fit the patient outlet easily and have minimum dead, space., 7 the system should withstand washing scrubbing and autoclaving procedures, 8 face masks : 3 sizes i.e. 00,01,02,: 3 set with each bag., 9 should have option to control peep & pip., si no. self inflating bag 750ml, 1 self inflating bag, 2 silicone made, 3 provided with closed ended reservoir with two valves, 4, patient valves pliable, well sealed, have minimum dead space and no forward or, backward leaks, 5, the bag should have an oxygen inlet which fits into the standard oxygen tubing both from, a cylinder and central supply, 6, face masks should be transparent, fit the patient outlet easily and have minimum dead, space., 7 the system should withstand washing scrubbing and autoclaving procedures, 8 face masks : 3 sizes i.e. 00,01,02,: 3 set with each bag., 9 should have option to control peep & pip., si no. laryngoscope with different size blades-, 1 high quality corrosion resistant stainless steel blades and body, 2 light source firmly fixed with blade, 3 blades size 00,0 and 1,2,3 sets with each, 4 straight blade for neonatal use, 5 should withstand chemical sterilization and autoclaving, 6 should be having fiber optic cold light source ., 7 should have xenon light illumination of more than 50,000 lux., 8 battery should be rechargeable for more than 2 hr. -quantity - 10 -msme exemption - yes | startup exemption - yes,
Tender For supply of laparoscopic telescope 5mm 30 degreetechnical specifications for a 5mm 30-degree laparoscope:general specifications diameter: 5mm viewing angle: 30 degrees length: typically around 300mm (can vary) material: stainless steel or titanium (for durability and bio compatibility) optics: high-quality glass lenses for clear image transmission light source: compatible with cold light sources (led) sterilization: autoclavable for repeated useimage quality resolution: high resolution for detailed visualization of surgical sites color accuracy: accurate color reproduction for proper tissue identification depth perception: good depth perception for precise surgical maneuversfeatures and accessories distal tip: designed for easy insertion and maneuverability objective lens: high-quality lens for optimal image quality light guide: efficiently transmits light from the source to the objective lens sheath: protects the scope during insertion and handling connectors: compatible with standard laparoscopic camera systemsadditional considerations compatibility: ensure compatibility with the surgeon's preferred camera system and light source.( we have a karl storz system, hence compatibility to be ensured) durability: the laparoscope should be durable and withstand repeated sterilization and use. ergonomics: a comfortable grip and design can reduce fatigue during long procedures.
Tender For supply of biofermenter in situ sterilization with accessories for working configuration 50 l biofermenter , 5 l seed fermenter , plc based centralized control panel , ph sensors , software supervisory control and data acquisition system for realtime bioprocessing control and monitoring of both 5l and 50l fermenters , laptop for scada software , steam generator electric boiler 18 kw , air compressor 10 cfm 100 l storage tank , chiller circulator with water circulation pump , ribbon mixer blender total capacity 300 kg , steam sterilizer chamber total capacity 50 kg , ss pipe line and fitting from fermenter to utilities , ultra pure water purification unit , do2 sensors
Tender For srph346350-vacutainer needle with visibility of blood entry 21g for multi-sampl e collection should have the following specifications: 1). the designing of the needles sho uld be such that it can able to minimize the pre- analytical variability throughout the speci men collection and handling process to ensure reliable results and should be suitable fo r faster procedures. 2). the needles should be suitable for closed- system blood collectio n i.e drawing blood directly from the vein into vacuum collection tubes to prevent exposu re of atmosphere. 3). the needles should be suitable for multi- sample collection.4). needle s should be sterile, double-ended with a sleeve covered safety valve for use with vacutain er tube holder and vacuum blood collection tubes. 5). size of the needle should be 21g x 1 i nch (gauge: 21g, length: 1 inch./25 mm).6) needle design should provide instant confirmation of needle placement, through visual flashback indication.7). there should be a chamber or similar arrangement at the needle junction for easy visual confirmation of bl ood flow entry.8). needle should have low-angle bevel design and should be coated with si licone/ similar low friction lubricant coating to ensure gentle and smooth vein entry (for smooth painless pricking to the patients 9). the firm should provide certificate of complian ce report for every batch and material used for making the needle and sterilization of the needle should comply with iso standards.10). the product should be bis/iso/ce/us fda certif ied and should produce the certificate while quoting in the tender.11). the firm should sup ply suitable needle holders along with the needle supply and also should submit adequate sampling material for checking suitability & for validation, verification of productunit:no [quantity tolerance (+/-): 5 %age , item category : normal , total po value variation permitted: max 8 lacs ]